TITAN VERSA C on Service Cart with Gas Cylinder image TITAN VERSA C, front view TITAN VERSA C, angled view TITAN VERSA C, rear view TITAN VERSA C Menus image TITAN VERSA C Bar Graphs graphic TITAN VERSA C Graphs graphic

The TITAN VERSA C (Compact) with external, user-supplied pump, is an incredibly powerful and versatile helium leak detector based on an innovative and proven core technology. It is lightweight, mobile, and flexible. It is most commonly mounted on a maintenance cart or integrated into a production leak test system. Typical applications include production system integration, service cart mounted for maintenance environments, and field service applications.

Founded on a core leak detector module, our configurable design—together with its high sensitivity and pumping capacity—gives you the flexibility and performance your application requires. Whether you need a service and maintenance tool or a production leak testing instrument, the TITAN VERSA is the versatile solution to your leak detection needs. 

The TITAN VERSA Warranty, Service, and Support package is the best in the industry. 

  • Warranty 
    • Leak detector - 1 year
    • Ion source - 2 years
    • Calibrated leak standard - lifetime
  • Service
    • Global network of Authorized Factory Service facilities
    • Full range of service options including self-service maintenance kits and comprehensive service contracts
  • Preventive Maintenance (PM) Service Agreements - simple, flexible and tailored to your needs
    • 15% discount on PM Service contracts at time of purchase
    • 72-hour uptime  guarantee
    • 24/7 product support
    • Free calibrated leak standard exchange
    • Discounted labor rates for other services when in a service agreement

MULTIPLE PURCHASE & RENTAL OPTIONS

  • Standard purchase
  • 2-year lease-to-own
  • Annual rental*
  • Short-term rental*

* Save 50% on rentals when combined with PM or repair services

TECHNICAL SUPPORT

Unmatched technical support from engineers and technicians with over 50 years of combined experience brings you industry leading support, including:

Project Design Assistance: Use our Contact form or contact us directly at 800-465-1004

  • Pump: None
  • Evacuation Valve: Standard flow
  • Vent Valve: Standard flow
  • Supply Voltage: 100-120 V~
  • Plug Type: N. America
  • CONFIGURE your custom TITAN VERSA

Click Here to view Available accessories

Primary Pump
None
Pumping Speed Capacity
N/A
Dimensions (W x H x D)
559 x 404 x 305 mm | 22.0 x 15.9 x 12.0 in
Weight
25 kg | 55 lbs
Operating Temp - Vacuum Method
0 - 45 °C | 32 - 113 °F
Operating Temp - Sniffing Method
0 - 45 °C | 32 - 113 °F
Electrical Supply (50/60 Hz)
90 - 250 V~
Electrical Power (Max)
350 W
Detectable Gases
Helium 4 | Helium 3 | Hydrogen
Min. Detectable Leak Rate -
Vacuum Method (Helium)
5 x 10-12 mbar L/sec | atm cc/sec
Min. Detectable Leak Rate -
Vacuum Method (Sniffing)
5 x 10-9 mbar L/sec | atm cc/sec
Max. Inlet Pressure - Vacuum Test Modes
GROSS - 25 mbar | FINE - 5 mbar | ULTRA - 0.5 mbar
(GROSS - 19 Torr | FINE - 4 Torr | ULTRA - 0.4 Torr)
Test Methods
Vacuum (includes AutoTest) | Sniffing
Helium Pumping Speed
2.5 L/sec (ULTRA mode)
Protection Category
IP 20
Inlet Flange
DN 25 ISO-KF
Start-up Time
≤ 3 minutes
Accessory Interfaces
USB port | Barcode | Accessory 1&2 | Wireless remote control | Wired remote screen
Communication Options
Serial ports: RS-232 | RS-485 | Ethernet
I/O Options
Remote I/O | Local I/O


Manuals / User Guides

TITAN VERSA High-Flow Valves User Manual
TITAN VERSA Quick Start Guide
TITAN VERSA Communication Interface User Manual
TITAN VERSA Lifting Guide
TITAN VERSA Helium Leak Detector User Manual
TITAN VERSA Remote I/O Module User Manual
TITAN VERSA Maintenance Cart User Manual
TITAN VERSA Helium Leak Detector User Manual - Spanish

Product Brochures

TITAN VERSA Product Overview
TITAN VERSA PM Services

Drawings

TITAN VERSA C Drawing
TITAN VERSA Maintenance Cart Drawing

Videos

TITAN VERSA Helium Leak Detector

Forms

Helium Leak Detector Rental Form

Reference Materials

Presentation: Production Leak Testing: Optimizing Helium Usage and Alternate Test Methods