Single-use bioprocess bags demand rigorous integrity testing to guarantee compliance with the highest manufacturing standards to safeguard the products processed and shipped in these containers. For effective leak detection and prevention of microbial ingression, the industry has widely embraced helium leak detection as the most sensitive method. For many single-use products, the accepted industry standard sets a maximum allowable leak limit (MALL) of a 2-micron equivalent hole size.
In 2012, LACO took the lead in pioneering the development of integrity leak test protocols and methods specifically designed to reliably detect defects in single-use bags.
Our cutting-edge technology has been successfully deployed at numerous leading manufacturing facilities worldwide. We are committed to upholding the highest quality standards and providing reliable solutions.
LACO’s unique standard platform single-use bag leak test systems include these key features:
Single-chamber or Dual-chamber systems
Choose between a single-chamber or dual-chamber system to meet your throughput requirements.
Allen Bradley or Siemens PLC
Choose which controller will be the best for your application.
| System Weight |
2200 lbs. |
| System Dimensions |
62″ W X 69″ H X 64″ D |
| Electrical Requirements |
220 VAC, 30Amps, 60hz, 3ph |
| Air Supply Requirements |
80-100 psi |
| Helium Gas Supply |
30 psi or above |
Perform fast, safe, and easy automated testing with the Allen Bradley PLC & HMI and included controls: