Custom Solution | TITANTEST Leak Test System for Bioprocess Containers

SKU:

X-1P29156

A worldwide manufacturer of scientific and medical products required a helium accumulation test for bioprocess containers system for leak testing disposable bioprocess containers. The system needed to detect very low leak rates across six part models while meeting aggressive cycle time requirements and providing full pass/fail and leak-rate data logging.

We engineered a custom automated leak testing system with dual clam shell-type aluminum test chambers with a sniffer probe to identify leaks at sealed ports. The system incorporates an automated sniffing accumulation test method and is connected to a helium mass spectrometer leak detector in order to achieve high sensitivity, test reliability, uptime, and fast cycle times.

  • Helium tracer gas
  • Dual clamshell-type aluminum test chambers
  • Automated sniffing accumulation test method
  • Fast cycle times
  • High sensativity
Application: Leak testing bioprocess containers
Industry: Medical & Pharmaceutical
Production Volume: Medium to high volume
Tracer gas: Helium
Controller: AB PLC
Leak Test Method: Helium accumulation
Test Pressure: 2 psig
Throughput: Four parts per minute
Cycle Time: 29 seconds
Configuration: Portable Workstation
Chamber: Two clam shell-style chambers
Leak Detector: TITANTEST P-Series Helium Leak Detector (discontinued)

 

Related Products