ATLAS Leak Test Process Controller – Integrated Leak Test, High Pressure

SKU:

ATD-4MDXHX-1

This ATLAS Leak Test Process Controller configuration enables automated leak testing, including part charging to 1000 psig. LACO’s TITAN VERSA helium leak detector is primarily integrated with this configuration, however, other leak detectors can be accommodated. The ATLAS Controller allows for a fully automated leak test process that is fast, repeatable, and reliable. Custom options available.

  • Fully integrated leak test controller – part fill during leak test process
  • TITAN VERSA leak detector integration
  • Fill pressures up to 1000 psig
  • Fully automated controller with flexible process control options
  • Flexible for the following test leak test configurations: helium bombing, hard vacuum, HATS, sniffing, accumulation
  • Software allows for fully automated tests along with system calibrations – up to 100 configurable test recipes
  • Test ID data entry: manually via touchscreen or via optional bar code reader
  • Controller includes spare I/O (digital and analog) to provide customized solution if needed

Custom Options

  • Custom data logging or alerts:
    • SQL database integration
    • OPC server integration
    • FPT server integration
    • Email alerts
  • Helium recovery valve and control
  • Flow/occlusion test
  • Custom part evacuation, fill, or vent sequence
  • Custom software or hardware to meet your needs
ATLAS Control Mode Pre-Charge
Leak Detector Integration Yes, TITAN VERSA or others
Gas Charge Pressure Range 0-135 psia (0-150 psia)
Regulator Control Manual
Part Evac Pump Diaphragm dry pump, 1.8 m3/h (1.1 cfm)
Compressed Air Gas 80 ““ 120 psig
Helium Gas 25 psig over max helium fill pressure
Main Enclosure Dimensions 16″ H x 20″ W x 6″ D
Main Enclosure Mounting Bottom or rear mounting
Main Enclosure Weight 35 lbs
Atlas Safety Option None
Atlas Tooling Option None
ATLAS HMI Configuration Remote HMI 7″ touchscreen
ATLAS HMI Dimensions 6.7″ H x 9″ W x 2.5″ D
ATLAS HMI Weight 2 lbs
ATLAS HMI Mounting VESA 100 mounting
Communication Protocols Ethernet, RS232 remote control, USB port
Communication Ports Ethernet x3 | USB 2.0 x2 | DB-15 – Remote IO x1
Remote I/O Port (DB-15) 24 VDC logic with 2 digital inputs and 4 digital outputs
Data Logging Memory 32 GB internal SD card, Data extraction via USB drive
Electrical 120 VAC | 50/60 Hz | single phase | 1000 watts max
Power Cord Nema 5-15P x IEC C13
Warranty 1 year

ATLAS Status Enclosure

The ATLAS Status Enclosure combines Start/Stop cycle control and Pass/Fail indicator lights in a single, easy-to-integrate accessory. Designed for production leak testing environments, it provides clear visual feedback and intuitive operator control during test sequences. The unit connects directly to...

ATLAS Ticket Printer

The ATLAS Ticket Printer provides fast, convenient printing of production leak test results for traceability and documentation. Once enabled in the ATLAS Controller software, it automatically prints test data on standard receipt paper, making it easy to log or archive...

ATLAS USB Wireless Barcode Reader

The ATLAS USB Wireless Barcode Reader enables fast and reliable data entry for any 1D or 2D barcodes. Designed for use with the ATLAS Controller, the barcode reader connects via USB and includes a wireless scanner and a USB charging...

USB Flash Drive, 32 GB

The 32 GB USB Flash Drive supports data logging and remote software updates on LACO leak detectors and controllers, including the TITAN VERSA, ATLAS, and LHCT. Ideal for transferring test records, calibration files, and firmware, this flash drive ensures seamless...

32 GB Micro SD Card

The 32 GB Micro SD Card enables reliable data logging and remote software updates for LACO leak detectors and controllers, including the TITAN VERSA, ATLAS, and LHCT systems. Designed for compatibility and convenience, it provides essential storage capacity for production...

Helium Bottle Regulator 0-100 psi

This helium regulator is designed for use in a variety of helium leak testing applications, including with spray probes, tracer gas manifolds, and sniff test setups. It supports a wide inlet pressure range and offers precise control for downstream components....

Air Filter Cartridge

This replacement Air Filter Cartridge is designed specifically for use with LACO’s ATLAS Controller. With a 5µm filtration rating, it helps prevent particulate contamination and ensures clean, consistent airflow through the system. Regular filter changes protect internal components and support...